德国ALV公司出品的ALV-LinCorr是一款高速(下限25ns)、实时线性数字相关器,并能支持以下模式:
★ 双通道模式
    两个独立的相关功能(独立的自相关或互相关),每个通道包含1024个线性通道+32个延迟通道。
★ 单通道实时模式
    一个相关功能(自相关或互相关),采样时间取决于相关通道数目,快达实时25ns,多至2048个相关通道。
★ 单通道批处理模式
    一个相关功能(自相关或互相关),2048个线性相关通道,在批处理模式下,所有采样时间低于250ns。通过数据批处理获得相关函数。直至整组实验完成,采样结束。
    取样停止,直到全部块处理完成。可达到的效率具有采样时间依赖性,当采样时间为237.5ns时,可达近75%的效率,当采样时间为下限值25ns时,效率略大于10%。
★ 直接通道的预延时
    对于实时或者批处理实验,直接通道都可以预延时至0 ~ 15875µs的范围(步进12.5ns),这就意味着可以对直接通道和延迟通道的传播时间进行补偿,以及在直接通道引入预延迟,允许使用“负滞后时间”的概念。
    所有现有ALV-LinCorr用户请注意,不管是批处理模式还是预延时的功能特点,都可以通过在线更新一个特殊的固件来实现。
    4×TTL@50欧姆计数输入,高达160 MHz的输入计数率。
    EPP和PCI接口,内置实验发生器,兼容windows-NT / 2000  /2003 / XP / Vista /Win7 / WIN10等系统。

技术 · Technology

    ALV-LinCorr是一款高速(25ns)、实时线性数字相关器,可达2048个相关通道(上限),支持自相关和互相关两种相关模式。该产品主要用于涉及周期相关功能的高速相关应用,如表面散射、调制外差探测、LDA、LDV等。

采用线性tau计算

ALV-LinCorr线性数字相关器